Technical Specifications
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 127
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P7500 is intended for use in the electromagnetic environment specified below. The customer or the user of the
P7500 should make sure it is used in such an environment.
Immunity Test
EN and IEC
60601-1-2
Edition 4.1
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Conducted
RF-Immunity
IEC 61000-4-6
Radiated
RF-Immunity
IEC 61000-4-3
3 V (80% AM)
150 kHz to
80 MHz
(6V in ISM
Bands per
CISPR-11)
3 V/m
80 MHz to
2700 MHz
3 V (80% AM)
150 kHz to
80 MHz
(6V in ISM
Bands per
CISPR-11)
10 V/m
80 MHz to
2700 MHz
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compli-
ance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with this symbol.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note 3: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the P7500 is used exceeds the applicable
RF compliance level above, the P7500 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the P7500.
IMMUNITY to Proximity Fields from Radio Frequency Wireless Communications Equipment
In addition to the Radiated RF IEC 61000-4-3 as shown in the table above, the P7500 has been tested as specified in the
table below.
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation Maximum
Power (W)
Distance (m) Immunity Test
Level (V/m)
385 380-390 TETRA 400 Pulse
modulation
18 Hz
1,8 0,3 27
450 430-470 GMRS 460,
FRS460
FM +
5 kHz devi-
ation
1 kHz sine
20,3 28
710 704-787 LTE Band 13, 17 Pulse
modulation 217
Hz
0,2 0,3 9
745
780
810 800-960 GSM 800/900
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
Pulse
modulation 18
Hz
20,3 28
870
930