Graphical Caregiver Interface (GCI) Controls
34 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
WARNING:
These warnings apply to both zero the scale and weigh procedures. To help prevent injury and/or
equipment damage, obey these warnings:
Warning—The scale is very sensitive. Make sure the bed is not touching anything that could
affect the patient weight zeroing or weighing process. Failure to do so could cause an
inaccurate weight reading.
Warning—Always use the New Patient zero before a patient is admitted into the bed. Failure
to do so may keep old patient data in the bed and cause a risk to the new patient.
Recommended Bed Position to Weigh a Patient and Required Bed Position to Zero the Bed
Head lower than 45° (A; head angle).
Foot not more than 30° below horizontal (B; foot up).
Trendelendburg/Reverse Trendelenburg less than 2°
(C; Trendelenburg angle).
Zero/New Patient
1. Make sure the patient is not on the bed.
2. Put the bed in the required position (refer to “Recommended Bed Position to Weigh a Patient and
Required Bed Position to Zero the Bed” on page 34).
3. Press the Scale menu control on the GCI screen.
4. Press Zero.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals