CPR Control
10 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
CPR CONTROL
The red CPR control pedals are located on each side of the base frame between the head-end and foot-
end casters.
USE THE CPR CONTROL
To Activate
WARNING:
WarningDo not use your hand to activate the CPR foot pedal. Injury could occur.
1. Step down and hold the red CPR pedal with your foot until the head section reaches the flat
position and you hear the audible tone. If you release the CPR pedal before the bed is flat, the head
section will stop.
The foot and knee sections will automatically move to a flat position from any position
including chair.
The surface automatically goes into Max-inflate/P-Max Inflate for 60 minutes. After 60
minutes the surface will go into Normal/Standard mode.
NOTE:
Use of a CPR board may increase the effectiveness of CPR.
2. To stop foot section movement, press any siderail control except Nurse Call.
Network connecting indica-
tor (flashing)—beds with
Bedside Association
Patient identity is not veri-
fied indicator. When flash-
ing, a patient is assigned,
but not verified, or the bed
has rejected the patient
identity—beds with Bed-
side Association
Network not connected indi-
cator—beds with Bedside
Association
When connected to AC power, the HandsFree CPR Control lowers
the head and knee sections, and raises the foot section. After the
head section is flat, a tone sounds and the foot section rises. The foot
section moves to a flat position within a maximum of 30 seconds if
fully articulated.
The integrated air surface will Max-Inflate/P-Max Inflate to provide a
firm surface to support a CPR board. After 60 minutes of Max-
Inflate/P-Max Inflate, the optional air surface will go into Normal
mode. If AC power is lost, the air surface stays at the level of pressure
that existed at the time of power loss.
Symbol Description Symbol Description

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals