Caregiver Siderail Controls
20 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
When you press and hold the Chair control, the bed will move through all of the chair positions.
Instructions will show on the GCI screen as the bed moves through the chair positions. Three beeps will
sound when the instructions show on the GCI screen.
DINING CHAIR Feature Positions
The Dining Chair feature allows the patient to be placed in a customized semi-seated position.
WARNING:
To help prevent injury and/or equipment damage, obey these warnings:
Warning—Do not transport a patient with the bed in a Dining Chair feature position.
Warning—Do not use ankle restraints when using this feature.
Warning—Observe lines, drainage bags, and linens closely during chair positioning.
FULLCHAIR Position
The FullChair feature is only available on beds with Chair Egress. It allows the caregiver to place the
patient in a fully seated position without having to remove the patient from the bed.
NOTE:
If the footboard is not installed, the bed will proceed into Chair Egress.
Dining Chair feature position, FullChair feature position, and Chair Egress
feature position with an Air Surface
To Activate
1. Make sure the brake is set.
2. Press and hold the Chair control. The patient deck
transitions to the reclined position (first the patient deck
will slightly recline backwards as the seat and lumbar
sections of the surface slightly deflate.) When the Chair
control is released, the seat section will re-inflate to normal
pressures (air surface beds only).
3. When the bed has reached the desired position, release
the Chair control. If desired, use the Head, Knee, Foot, or
Foot Retract controls to make custom Dining Chair
feature position adjustments.
To Activate
1. Set the brake.
2. Press and hold the Chair control. The patient deck
transitions to the reclined position (first the patient deck
will slightly recline backwards as the seat and lumbar
sections on the surface slightly deflate) then to the Chair
position.
3. If the footboard is installed, when the articulation stops
and a tone sounds, the bed has reached the FullChair
feature position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals