Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 139
P7500A BED UPGRADED WITH A P7540A PROGRESSA+ SURFACE
Warning—Do not use a Progressa+ Xtend (wide) surface with a Progressa Bed (P7500A). To do so
could cause patient injury or equipment damage.
Do not use a Progressa+ P7540 surface on any Progressa P7500A bed unless the Surface Upgrade Kit
(P7541A01, P7541A02, P7541A03, P7541A04, or P7541A05) has been completed and the compatible
P7540A surface has been selected from the Instructions provided with the upgrade kit.
The Progressa bed (P7500A) can be upgraded to support the Progressa+ Accelerate and Advance
surfaces.
Warning—Only narrow Progressa+ surfaces can be used on the Progressa Bed (P7500A) or patient
injury could occur.
Product Number Revision Options Digit Code
P7540 A XXXXXXX
Digit 7 =Xtend feature (40" surface width) 0:No/1:Yes
Digit 8 = StayInPlace feature 0:No/1:Yes
Digit 9 = Pulmonary Therapy Mode 0:no/1:CLRT only/
2:CLRT and P&V
Digit 10 = Alternating Low Pressure mode 0:No/1:Yes
Digit 11 = Rental option 0:No/R:Yes
Digit 12-13 = 0 (not used)
Accelerate Surfaces Advance Surfaces
Bed Frame
Compatible Surface
Product Numbers
(See the surface product label at
the head end of the
surface.)
Bed Frame
Compatible Surface
Product Numbers
(See the surface product label at
the head end of the
surface.)
Dining Chair
P7540A0010000
Dining Chair
P7540A0000000
P7540A0020000
P7540A0011000 P7540A0001000
P7540A0021000