Graphical Caregiver Interface (GCI) Controls
30 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
If you want Bed Exit set during a pulmonary therapy (Rotation or Percussion and Vibration), Bed Exit
must be initiated before starting the therapy. Only Out of Bed Mode will work during a pulmonary
therapy.
Turn OFF the Alert
1. Press the Alerts menu control on the GCI screen.
2. Press Bed Exit.
3. Press Off. This turns off the Bed Exit Alert.
Pre-Emptive Alert Silence
When the Bed Exit System is on, it can be silenced with the Pre-emptive alert silence control in the lower
left corner of the screen for 30 seconds and then suspended for 10 to 30 minutes without turning the
system off.
3. Press Bed Exit.
4. Press one of these:
• Position
•Exiting
•Out of Bed
Only one bed exit mode can be active at a time.
A message will show when the bed exit alert is active.
– When armed, the alert indicator will turn green on the home screen and the center of the
indicator will show the selected mode of sensitivity.
Patient Position Mode Exiting Mode Out of Bed Mode