Graphical Caregiver Interface (GCI) Controls
42 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
After the scale is zeroed/tared, and the empty bed is in a stable position, a green indicator with
+/- 0,25 e Zero/Tare will appear on the Scale screen. This indicates the bed has an acceptable zero/tare.
Once there is weight in the bed this indicator will not show. If there is an unstable equilibrium +/- 0,25 e
Zero/Tare indicator will also not show. If the empty bed has been zeroed/tared, is in a stable position,
and the indicator is not on, the bed will need to be re-zeroed/tared.
Magnification Mode (Extended Weighing Device)
4. Press Zero/Tare.
5. Press:
New Patient
Erases Scale History (all previously recorded
patient weights will be erased)
Zeroes the scale
Returns the surface to Normal /Normal (CLP)
mode
Turns off all RemindMe reminders
or
Zero/Tare
–Does not erase Scale History
Zeroes/Tares the scale
Only available on the NAWI Compliant (EN 45501)
Scale. Pressing the magnifying glass (Magnification
Mode (A)) will change the scale display increments
to 0.1 kg for 5 seconds. Weights cannot be saved in
the Magnification Mode.
Magnification Mode (A)—changes the
scale display increments to 0.1 kg for 5
seconds.
Bed not in recommended position
(B)—shows the recommended bed
position for weighing.
Unstable equilibrium indicator (C)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals