Class III
e= 0.5
Conforms to the European Medical Device Directive 93/42/EEC for a device that has a measuring
function. The scale is classified per Scale Directive 2009/23/EC.
Maximum weight: 250 kg
Minimum weight: 10 kg
Display interval: 0.5 kg
Combined zero and tare range: 10 kg to 250 kg
The maximum scale capacity is 250 kg, however the maximum patient weight for the bed is 227 kg.
2. Press Options.
3. Press Add/Remove Items. Follow the on-screen
instructions.
After using the Add/Remove function the word Net will
appear next to the non-verified weight. Net indicates a
user has manually changed the non-verified weight. The
weight saved after using Add/Remove Items will be noted
with the word Net next to it. If that same device is later
removed or the cumulated adjustment amount is 0 kg, the
word Net will no longer appear.