Preventive Maintenance
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 111
ATMOSPHERIC PRESSURE PRODUCT PERFORMANCE
The following surface functions may have reduced performance at higher altitudes:
• Percussion and Vibration Therapy—Less effective
• Other Inflation Functions—Slower achieving maximum level
PREVENTIVE MAINTENANCE
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—Only facility-authorized personnel should perform calibration on the Progressa
Bed.
• Warning—Only facility-authorized personnel should perform preventive maintenance on the
Hillrom Progressa ICU Hospital Bed.
The Hillrom Progressa ICU Hospital Bed requires an effective maintenance program. We recommend
that you perform annual preventive maintenance (PM) and testing for Joint Commission certification.
PM and testing not only meet Joint Commission requirements but can help support a long, operative
life for the Hillrom Progressa ICU Hospital Bed. PM will minimize downtime due to excessive wear. For
the preventive maintenance detailed procedures, refer to the Hillrom Progressa and Progressa+ ICU
Hospital Bed Service Manual (171748).
Perform annual preventive maintenance procedures to make sure all bed components are functioning
as originally designed. Pay particular attention to safety features, including but not limited to:
• Siderail latching mechanisms
• Siderail dampers for oil leaks
•Caster braking systems
• Electrical system components
• Electrical power cords for fraying, damage, and proper grounding
• All controls return to off or neutral position when released
• Controls or cabling entanglement in system mechanisms or siderails
• Proper operation of the lockout controls
• Integrity of sleep surface cover
• Intake/Exhaust filters for cleanliness and condition
Main Battery
Replace the battery if any of these conditions exist (refer to the Hillrom Progressa and Progressa+ ICU
Hospital Bed Service Manual (171748):
• The battery indicator does not come on within 3 minutes of bed connection to AC mains.
• The battery indicator does not increase the number of illuminated LEDs within 12 hours of bed
connection to AC mains.
IntelliDrive Transport System Batteries
Replace the batteries if the IntelliDrive Transport System automatically shuts down power before the
final battery charge indication LED flashes (refer to the Hillrom Progressa and Progressa+ ICU Hospital
Bed Service Manual (171748)).