Caregiver Siderail Controls
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 15
WARNING:
WarningLock out all articulation controls when traction equipment is installed. Failure to do so could
cause patient injury.
To Activate
At the same time press the Lockout control and the function control.
A single beep sounds and the locked function indicator will stay on. Both patient and
caregiver controls are locked out.
If the lock out procedure is done incorrectly, the bed will beep three times and a screen will
show on the GCI screen to show the correct procedure.
The knee lockout will lock out the foot control. The foot Up/Down lockout will lock out the
knee control.
The bed up down lockout will lock out Trendelenburg and Reverse Trendelenburg.
Any lockout will also lock out all chair positions and bed flat.
To Deactivate
Disable any lockout by simultaneously pressing the Lockout control and the respective function
control. A single beep will sound when the lockout is deactivated.
BED UP/DOWN
CAUTION:
Caution—Make sure there is sufficient headwall clearance when you raise/lower the bed. Equipment
damage could occur.
To Activate
Press and hold the Bed Up control to raise the bed or press and hold the Bed Down control to lower
the bed. Release the control when the desired height is reached.
To disable the Bed Up/Down - Activate the Up/Down lockout control.
NOTE:
While holding the Bed Down control, the bed motion will slow down just before reaching the lowest
position. Continue holding the Bed Down control until the bed stops completely. When the bed reaches
the lowest position, the “Bed Not in Low position indicator” on the intermediate siderail control panel
will go out and the bed position indicator on the home screen of the GCI screen will turn green.
The Bed Up/Down controls are located on the head-end siderails and on the intermediate
siderails. The controls adjust the height of the bed from a low position for patient exit to a
high position for examination. To lock out a control, refer to “Lockout Controls” on page 14.
WARNING:
WarningLowering the bed may cause linens, drainage bags, and other equipment to
come in contact with the floor. Follow facility protocol if they touch the floor.