Technical Specifications
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 125
Policies and procedures must be established to train and educate your staff on the risks associated with
electric equipment. It is never prudent or necessary for personnel to place any part of their body under
or between moving parts of the bed. Whenever a bed is being cleaned or serviced, it should be
unplugged from its power source, and the lockouts should be activated to keep the bed from
accidentally operating due to the battery backup. Refer to the Hillrom Progressa and Progressa+ ICU
Hospital Bed Service Manual (171748).
WARNING:
WarningThe P7500 should not be used adjacent to or stacked with other electrical equipment. If
adjacent or stacked use is necessary, observe the P7500 and the other electrical equipment to make
sure they operate as intended.
Make sure the P7500 operates correctly when it is used near other electronic devices. Portable and
mobile radio frequency (RF) communications equipment can affect electrical equipment.
Medical equipment needs special precautions in regard to electromagnetic compatibility (EMC) and
needs to be installed and put into service according to the EMC information supplied in the tables that
follow.
NOTE:
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is
normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or
reorienting the equipment.
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The P7500 is intended for use in the electromagnetic environment specified below. The customer or the user of
the model P7500 should make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1 The P7500 uses RF energy only for its internal func-
tions. Therefore, its RF emissions are low and are not
likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class A The P7500 is suitable for use in all establishments other
than domestic establishments and those directly con-
nected to the public low-voltage power supply net-
work that supplies buildings used for domestic
purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals