EasyManuals Logo

baxter Hillrom Progressa User Manual

baxter Hillrom Progressa
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
Caregiver Siderail Controls
22 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
4. For beds with a Progressa Prevention Surface with the Chair Egress feature, continue to press and
hold the chair control until the bed is at the chair egress position.
5. For beds with an air surface, the GCI screen will indicate when the seat is deflating. Wait until the
surface deflates fully and the bed beeps three times.
NOTE:
Pressing cancel on the GCI screen will re-inflate surface.
6. Press and hold the Chair control again. The frame will tilt forward to put the patient's feet closer to
the floor.
One beep will sound when maximum tilt is reached.
7. If needed, press and hold the Chair control to inflate the back section of the surface to the desired
amount to assist with the patient egress.
8. Make sure the patient's feet are on the floor and clear of all obstructions and trip hazards including
the deflated surface and linens. Monitor the patient and patient lines during bed egress. Assist the
patient with egress.
WARNING:
WarningWait for the bed to complete all frame articulations and surface deflations and the patient’s
feet are touching the floor before the patient exits the bed. Patient injury could occur.
WARNING:
To help prevent injury and/or equipment damage, obey these warnings:
Warning—The patient’s feet must be supported by the floor at all times while in the exit chair
position. Injury to the patient may occur from improper positioning.
Warning—Do not transport a patient with the bed in the chair egress position.
Warning—Do not use ankle restraints when activating this feature.
Warning—Observe lines, drainage bags, and linens closely during chair positioning.
Warning—If bed sheets contact the floor during Chair Egress, follow standard infection control
procedures.
Warning—Make sure the that sheets and blankets are not in a location where a patient will
step on them when they exit the bed. Bedding in a location where it can be stepped on
presents a trip hazard that can cause patient injury.
NOTE:
The patient deck first tilts back and then lowers as it goes to the FullChair feature position.
When the bed beeps three times, follow the on-screen prompts to help you through the
correct procedure for Chair Egress.
Monitor the patient, patient lines, and drainage devices.
For patient comfort, remove patient pillows before moving the bed to the chair egress
position.
For patient safety, remove the top sheet and any other items that may restrict leg movement
before egressing the patient from the bed.
To Deactivate
To move the bed out of a chair position, press and hold the Bed Flat control.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals