pro+ Surface Instruction for Use (209196 REV 8) 103
Specifications
IMMUNITY to Proximity Fields from Radio Frequency Wireless Communications
Equipment
In addition to the Radiated RF IEC 61000-4-3 as shown in the table above, the Surface has
been tested as specified in the table below.
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation
Maximum
Power
(W)
Distance
(m)
Immunity
Te st Leve l
(V/m)
385 380-
390
TETRA 400 50% PM,
18 Hz
1.8 0.3 27
450 430-
470
GMRS 460;
FRS 460
50% PM,
18 Hz
20.328
710 704-
787
LTE Band 13, 17 50% PM,
217 Hz
0.2 0.3 9
745
780
810 800-
960
GSM 800/900;
TETRA 800;
iDEN 820;
CDMA 850;
LTE Band 5
50% PM,
18 Hz
20.328
870
930
1720 1700-
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3, 4, 25;
UMTS
50% PM,
217 Hz
20.328
1845
1970
2450 2400-
2570
Bluetooth;
WLAN 802.11
b/g/n;
RFID 2450;
LTE Band 7
50% PM,
217 Hz
20.328
5240 5100-
5800
WLAN 802.11 a/n 50% PM,
217 Hz
0.2 0.3 9
5500
5785
NOTES:
For frequencies above 1 GHz, the transmitting antenna distance was reduced to 1m as
permitted by the IEC/EN60601-1-2 and UEC/EN61000-4-3 Standards
As an alternative to FM modulation, 50% pulse modulation at 18 Hz mat be used
because while it does not represent actual modulation, it would be worst case as
permitted by the standard (Table 9 of IEC 60601-1-2: 2014).