pro+ Surface Instruction for Use (209196 REV 8) 55
VERSACARE Bed—pro+ Surface Installation Instructions
Warning—Do not allow the surface to stay in continuous contact
with the headboard. This could impact the scale accuracy and Bed
Exit performance which could cause patient injury.
36. Make sure the surface is not in continuous contact with the
headboard.
37. If the bed has a scale system, zero the scale. See the bed’s user
manual.
38. If the bed has the Obstacle Detect System, raise and lower the bed
to make sure the surface power cord does not interfere with the
system. See the bed’s user manual.
Warning—Put the power cord in a location where persons will not
trip over it, and away from bed mechanisms. Failure to do so could
cause injury or equipment damage.
39. Make sure the power cord is in a location where persons will not
trip over it, and away from bed mechanisms.