Number of counts of main photodiode Minimum Maximum
Wavelength 560 550000 950000
Wavelength 600 750000 950000
Wavelength 635 850000 950000
Wavelength 670 450000 950000
Number of counts of reference
photodiode
Minimum Maximum
Wavelength 340 100000 400000
Wavelength 405 400000 800000
Wavelength 505 400000 800000
Wavelength 535 400000 800000
Wavelength 560 400000 950000
Wavelength 600 300000 800000
Wavelength 635 400000 800000
Wavelength 670 200000 950000
Measurements taken at the
505 nm wavelength
Parameter Value
Repeatability measured at
0 Abs
Standard
deviation
< 0.0002 Abs
Stability at 0 Abs Range < 0.0015 Abs
127