EasyManua.ls Logo

biosystems BA 400 - Metrology

biosystems BA 400
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.5.2.2. Metrology
is option is used to check the analyser photometry status. See Figure 6.15.
e following checks can be made:
Repeatabilty of the readings.
Stabiliity of the readings.
Absorbance reading.
Figure 6.17 Photometric verication screen.
To make a verication the rotor must be lled with distilled water. is can be performed by the analyser or the
user can do it manually.
1. Select the rotor lling mode and the well where the readings will be made
2. Press the icon for raising or lowering the wash station in order to remove or insert a rotor.
3. If you want to measure the absorbance of a specic reagent or uid, select the manuall lling option. Dispense
the uid to be measured into a well. In the screen, select the well where the uid was dispensed.
4. Select the type of measurement to be made.
5. Press the adjustment button to start the measuring process.
Repeatability
is verication takes measurements in all the wavelengths for 3 minutes. It is used to verify the repeatability of
the photometric system readings
After taking the measurements it shows the following parameters for each wavelength:
Mean
Standard deviation of all measurements
Coecient of variation (CV)
Maximim absorbance
Minimum absorbance
See chapter AIII ranges accepted for parameters.
79

Table of Contents

Related product manuals