EasyManua.ls Logo

biosystems BA 400 - Ise Module

biosystems BA 400
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 6.27 Barcode reader adjustment screen
6.5.7. ISE module
is menu is used to launch the dierent maintenance actions required for the ISE module.
e following functions can be executed:
Calibrate
Install a reagent kit
Install the electrodes
Deactivate the module for a long period of time
Change the peristaltic pump tubes
Activate the ISE preparations
For each function several actions must be executed. Select one of the functions and show the set of actions to be
executed step by step.
See the explanation of each step in detail in chapter14.2.2 of the user manual.
Select an action and press the execute button. Information about the action will appear in the results zone. It will
say whether the action was successfully completed (the text is shown in black) or has errors (the text is shown in
red). e results are shown in the actions that return information, such as calibrations.
In addition each of the actions is positioned in a group under the name General; if the user only wants to perform
one of the actions, that user can launch it directly.
Action Description
Maintenance Empties the tubes. Only activates the waste pump. In
the repetitions parameter indicate how many times the
action must be executed.
Bleed A Performs a priming cycle with calibration standard A
using a volume of 100 μL. In the repetitions parameter
indicate how many times the action must be executed.
Bleed B Performs a priming cycle with calibration standard B
using a volume of 100 μL. In the repetitions parameter
indicate how many times the action must be executed.
87

Table of Contents

Related product manuals