AII. Technical characteristics
e manufacturer declines all liability for damages caused by using the
appliance incorrectly.
GENERAL CHARACTERISTICS
Speed 400prep/h (without electrolytes)
ISE module speed 320prep/h
Analysis principles Colorimetry, turbidimetry.
ISE module: Potentiometry (selective electrode
method): Na
+
, K
+
, Cl
-
(Li
+
is optional)
SAMPLE CONTROL
Sample rotor capacity 135
Barcode detector Yes
Number of samples with barcodes 90
Size of primary tubes Diameter 12 mm to 16 mm (max. height 100 mm)
Sample well Sample well diameter 13.5 mm
Type of sample pump syringe Low-maintenance ceramic piston
Piston diameter 3 mm
Pipetting volume 2 μL to 40 μL
Pipetting resolution 0.1 μL
Maximum ratio between sample and reagent volume 1:2 to 1:200
Level detection Yes
Washing of tip Interior and exterior
Clot sensor Yes
Vertical collision detector Yes
REAGENT CONTROL
Reagent bottle volume 20mL, 60mL
Reagent rotor capacity 88 (44 bottles of 20mL or
60mL + 44 bottles of 20mL)
Refrigerated reagents Yes
Maximum refrigerator temperature 5° C to 8° C (at room temperature of 25° C)
Barcode detector Yes
Reagent arms 2 (R1, R2)
R1 reagent volume
150 mL to 500 mL
R2 reagent volume
40 mL to 300 mL
Type of reagent pump syringe Low-maintenance ceramic piston
Piston diameter 8 mm
123