EasyManua.ls Logo

biosystems BA 400 - Reaction Rotor; Optical System; Wash Station

biosystems BA 400
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Both 60 mL and 20 mL bottles can be placed in the inner ring.
e rotor is refrigerated, and the mean temperature inside it is below 8degrees C. e fridge has an independent
power supply system with its own switch so that when the analyser is switched o, the fridge can continue to operate.
2.3. Reaction rotor
e reaction rotor has 120 positions and a cover. e whole assembly is refrigerated.
e rotor is a single part made of PMMA and has 120 positions. e sample and reagent mixture reactions are
performed on it. Dierent optical readings are taken during the reaction. e PMMA material lters UV radiation.
e rotor is maintained at a stable temperature of 37 degrees C by a Peltier-based thermostatting system.
e reaction volumes range is from 200 μL to 600 μL.
Each arm dispenses in a dierent position from the rotor. e dispensing positions are:
Cycle 1: R1 dispensation
Cycle 31: S dispensation (sample)
Cycle 32: R1+S stirring
Cycle 33: Initiation of photometric readings
Cycle 66: R2 dispensation and stirring of reagent 2
Cycle 100: End of the reading processes
Cycles 101 –111: Elimination of the completed reaction and cleaning of the sample well in the wash station.
2.4. Optical system
e optical system is located in the reaction rotor, below the wash station.
It is formed by a set of leds, lters, beam splitters, the reaction rotor and two photodiodes.
ere is one principal reading diode and a reference photodiode that enables the correction of perturbations gener-
ated in the light source
e analyser has 8 wavelengths: 340, 405, 505, 535, 560, 600, 635, 670
e measuring range is from -0.2Abs to 3.5 Abs.
e measuring resolution is 0.0001 Abs.
e system automatically performs a blank on the cuvette before dispensing the reagent. is blank absorbance
value is used to correct the absorbance values measured in the cuvette.
If this value exceeds a preestablished limit, the cuvette is discarded.
2.5. Wash station
e wash station consists of an assembly with 7 tips located above the reaction rotor.
Each tip has a specic function and corresponds to a dierent execution cycle:
Cycle 1: Suction of the high contamination waste and dispensing of the washing solution
Cycle 2: Suctioning and dispensing of the washing solution
Cycle 3: Washing solution rest phase

Table of Contents

Related product manuals