Figure 6.18 Scales adjustment screen
6.5.3.2. Internal bottle verification
Detecting the level of the internal bottles containing distilled water and low contamination waste is done through
a buoy system. is means that only the analyser can detect whether a bottle is full or empty.
e programme enables the status of the buoys to be checked and indicates whether they are full or empty.
e status of the buoys for each bottle is shown on the screen (see Figure 6.17). e verication may be made
manually or automatically.
Manual verification
Access the buoys of the internal bottles. To do this, access the internal bottles through the rear part of the analyser
and unscrew their caps.
Move each buoy upward and downward. e status of each buoy will be constantly shown on the screen.
Automatic verification
Press the Start button for the programme to perform a full lling and emptying cycle on each bottle. e programme
will constantly indicate the status of the buoys.
Bear in mind that this automatic verication process may take a few minutes.
81