54
Test point Function
TP11 24 V Fridge
TP12 5 V Fridge
TP13 CAN_L
TP14 Board selection. Direction 4
TP15 Board selection. Direction 3
TP19 Rotor stepper motor
LEDs Function (on condition)
LD1 3.3 V voltage
LD2 3.3 V voltage Fridge
LD3 Barcode reader activation
LD4 Microprocessor start-up
LD5 Cover detection
LD7 Rotor start-up detection
LD9 Fridge peltier activation
LD10 Fridge peltier activation
LD12 Fridge microprocessor start-up
4.8. Fluid control board – CIIM00053
is board is located at the top of the uid zone.
e board controls the following elements, through a micro controller (U10):
• Stirrer driver (U18)
• Scales signal conditioning (U5 and U6) for the washing solution and high contamination waste
• Internal water and waste bottle buoy conditioning
• Driver for the wash station water heater (U1)
• Temperature sensor for controlling the thermostatting of the wash station water
• Wash station water dispensing driver
• Wash station water dispensing start-up detection