REPACKAGING RECORD
Company name:
Serial number:
Repackaging date:
PART REPLACEMENT PROCESS
1. Parts to be replaced
Replacement of parts Ok / Not Ok
Update the rmware
and software
Ok / Not Ok
ADJUSTMENT PROCESS
Points to be readjusted Tolerances Value
1. ermostatting of reagent 1 tip
Temperature set point 50,0 55,0
Temperature test 30,0 35,0
2. ermostatting of reagent 2 tip
Temperature set point 50,0 55,0
Temperature test 30,0 35,0
3. ermostatting of reaction rotor
Temperature set point 37,0 39,0
Temperature test 36,8 37,2
4. Photometry Tests
4.1. Luminous intensity
340 nm
6000 25000
405 nm
14000 30000
505 nm
12000 30000
535 nm
4000 30000
560 nm
3000 15000
600 nm
6000 25000
635 nm
6000 20000
670 nm
4000 15000
4.2. Number of counts of main photodiode
340 nm
600000 950000
405 nm
850000 950000
505 nm
850000 950000
535 nm
850000 950000
560 nm
550000 950000
600 nm
850000 950000
635 nm
850000 950000
670 nm
450000 950000
4.3. Number of counts of reference photodiode
340 nm
100000 400000
405 nm
400000 800000
505 nm
400000 800000
535 nm
400000 800000
560 nm
400000 950000
600 nm
300000 800000
635 nm
400000 800000
670 nm
200000 950000
4.4. Dark current counts
Value 3300 4500
4.5. Repeatability Tolerances CV(%)
For all the lters < 10%
4.6. Stability
For all the lters < 10%
FINAL FUNCTIONAL CHECK
5. Biochemical precision
Tolerances CV(%)
Value
Control serum Level I
Glucose
< 2
Control serum Level II
Glucose
< 2
Control serum level I
ALT
< 4
Control serum Level
II ALT
< 2
FINALISATION
6. Completed
Label the renovated
equipment
Ok / Not Ok
Add a full box of acces-
sories
Ok / Not Ok
Check the packaging Ok / Not Ok
7. Acceptance of repackaging
Repackaging OK
Available for sale
Ok / Not Ok
Signed:
Prepared by:
Page 1/1 Rev 1.0 Operative date: 25-06-12