Appendix B B-19
An aborted read is indicated by a <DLE> response at any time. Any data received,
including checksums, should be considered invalid and irrelevant in this case.
After data for the very last plate has been sent, the reader performs some final FMEA
tests to ensure that no errors may have occurred during the read process. If any
errors were detected a <DLE> character will be transmitted as a final series
terminator (the actual error code will be shown on the reader display, where the user
will have to go to stop the beeping). Otherwise another <^Z> character will be sent.
This indicates that the reader is completely finished with the read, and that more
commands can be sent by the host.
Since the command function described above is new, there is no reason to maintain
compatibility with obsolete products. Therefore, no assay name, report title, or
time/date stamp - blank or otherwise - will be returned with data collected using the
above function.
FILTER CONFIGURATION
To select wavelengths to be used in future read cycles, use the currently defined Set
Filter Table command ('M'):
Host Limits Response
‘M’ <ACK/NAK>
first wavelength (3) ‘200’ - ‘800’
comma separator (1) ‘,’
second wavelength(3) ‘000’ - ‘800’
comma separator (1) ‘,’
last wavelength (3) ‘000’ - ‘800’
comma separator (1) ‘,’
<ETX>