EasyManua.ls Logo

CETAC ASXPRESS PLUS - Replacing or Reorienting the 6-Port Valve

CETAC ASXPRESS PLUS
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ASXPRESS PLUS Rapid Sample Introduction System Operator’s Manual
Chapter 11: Maintaining the ASXpress Plus Rapid Sample Introduction System
121
Replacing or Reorienting the 6-Port Valve
The 6-Port valve assembly has a finite lifespan that is dependent upon the
conditions and sample media to which it is exposed. Exposure to higher
sample solids levels reduces the valve lifespan.
To determine whether the 6-port valve requires replacement, inspect the unit
for these conditions:
Valve dripping or leaking from the overflow hole behind port #4 at bottom
of the valve body.
With no other apparent problems, air is present in the lines (indicating a
leak or poor seal).
The valve can also be reoriented so that the nebulizer port is as close as
possible to the nebulizer.
Note that any time the 6-port valve body is removed from its actuator, the valve
will require retraining (reinitialization).
To replace or reorient the 6-port valve:
1 Unplug the valve/pump module cable from the electronics module.
Leave the electronics module powered on.
2 Remove the tubing/connectors on the 6-port valve.
3 Using the provided hex key, loosen the hex screw on the locking collar which
secures the base of the valve to the body of the valve/pump module.
Figure 11-5 Locking collar with hex screw loosened
4 Place the pump/valve module on its back, on a tabletop.
5 Firmly but carefully pull the valve straight up.
Now, before you reinstall the valve, retrain the valve actuator:
6 Start the Xpress Configuration Tool and make sure it is communicating with
the electronics module.
7 Expand the GUI using the arrow button on the right side of the window.

Table of Contents

Other manuals for CETAC ASXPRESS PLUS

Related product manuals