Danatronics, Corp.
100
Range: Displays how much material range, e.g. 5.000"
Delay: Delays start of signal
Main Bang Blank (MB blank): Used to eliminate main bang recover noise
ECHO Window: End of maximum time (material) to be measured
ECHO 1 Detect: + or –
Pulser Voltage: Pulsr volt: 30, 60, 90, 120V
ID: If datalogger equipped to change to another ID location
Alarm: High, Low, High-Low, Off
If another type of probe is chosen (e.g., delay line), other fields such as IF blank, C3
blank, and echo 2 polarity will appear as additional setup parameters.
Note, if any parameter has been adjusted, then an asterisk will appear next to the probe
setup file. You have a choice at this point to Save above F2 and give it a file name or
Discard above F3. You can save up to 2,700 unique application setups.
For all other gage setups such as alarm, hold, datalogger, etc. please refer to the ECHO 9
portion of this manual.
ECHO 8 allows you to choose between corrosion and precision gage functionality from
the Menu list, under Initial Settings, followed by the Gage Type parameter: