ERROR CODES AND RESOLUTIONS
PRODUCT REFERENCE GUIDE
92
207
In Beam Sensor
Error
Error
The scanner has
an in-the-beam
sensor, which
provides timing
for the decoding
of barcodes. The
scanner moni-
tors this signal,
and posts this
error if the signal
is sporadic or
missing.
The scanner
stops reading
barcodes.
Put the scanner in
the test mode to
test its ability to
read the barcode
(see “Diagnostics |
Read Test” on page
186).
This is an inter-
nal failure and is
not field ser-
vice- able.
Replace scanner.
211 PTP Error Error
Cannot synchro-
nize the internal
clock using the
PTP (precision
time protocol)
protocol. This
clock sync pulse
is generated by
the controller
unit. It is used to
sync the read
data of all the
scanners in the
network with the
master unit.
LEDs flashing.
Confirm that all the
scanners are con-
nected in the chain.
Use the scanner
test mode to con-
firm that the scan-
ner can statically
reading a barcode
(see “Diagnostics |
Read Test” on page
186).
Position a bar-
code on a box so it
faces the question-
able scanner, and
dynamically test
the scanner.
Replace faulty
scanner.
219
Fieldbus Comm
Error
Error
Cannot commu-
nicate with the
fieldbus module.
- The host loses
communication
with the scanner.
Confirm that the
unit is not commu-
nicating with the
Host.
Confirm that the
fieldbus parame-
ters are configured
properly (see
“Fieldbus” on page
122).
- Replace field-
bus module.