Alarms
192 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Enabling CO2 alarms:
1. Press the >CO
2 Alrm OFFon< key
(1 in Figure 116).
Disabled CO
2 alarms are enabled automatically
when changing to another ventilation mode.
The alarms for etCO
2 and FiCO2 can be
activated or deactivated in Standby for switching to
Man/Spont
When the alarm limits are enabled the value is
adopted from the automatic ventilation mode, see
page 209.
Figure 117. Location of >Config< KeyCO2 alarms can also be enabled and disabled
globally for all ventilation modes:
1. Press the >Config< key (1 in Figure 117) in the
standard or data screen.
The submenu Volumes/Alarms is opened (refer
to Figure 118).
2. Select and confirm the column “Alarms On/Off”
via the rotary knob.
3. Select and confirm the line “CO
2” via the rotary
knob.
4. Select and confirm “On” or “Off” via the rotary
knob.
CAUTION!
Risk of inadequate monitoring
National standards require a minimum monitoring
with some alarm functions. These standards may
not be met if the alarm function of the etCO
2
monitoring parameter is disabled.
Only disable this monitoring parameter after
consulting national standards.