Troubleshooting
268 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Advisory
(10)
AIR PIPELINE FAIL Compressed Air supply has failed. Open optional backup Air cylinder.
Check piped medical Air supply.
Pipeline supply hose not connected or
kinked.
Check connection to piped medical Air
supply.
Optional Air cylinder is empty or closed. Connect a full Air cylinder or open the
cylinder valve.
Compressed Air compressor has failed. Check compressor.
Caution
(24)
Warning
(31)
APNEA Priority in accordance with maximum
priority of the individual alarms.
Breathing/ventilation has stopped
(detected by pressure, volume, and CO
2
monitoring).
Patient must immediately be
ventilated manually!
Check patient's spontaneous breathing
ability.
Check ventilator settings.
Check fresh-gas setting.
Maske sure everything is connected.
Check hose system and tube.
Advisory
(10)
Caution
(24)
Warning
(31)
APNEA CO
2 Apnea alarms are graded in time (see
page 210).
In automatic ventilation modes:
Caution = 0to30sec.
Warning = >30 sec.
In ventilation modes Man/Spont,
Pressure Support, Aux CGO:
Advisory = 0to30sec.
Caution = 31 to 60 sec.
Warning = >60 sec.
Sample line not connected. Check sample line.
No spontaneous breathing. Patient must immediately be
ventilated manually!
Check patient's spontaneous breathing
ability.
Make sure everything is connected.
Check hose system and tube.
Breathing/ventilation has stopped. Patient must immediately be
ventilated manually!
Check ventilator setting.
Priority Message Cause Remedy