Troubleshooting
Instructions for Use Apollo SW 4.5n 269
Part Number: 9053586, 3rd edition
Caution
(see
page 213)
/
Advisory
(see
page 222)
(10)
Caution
(24)
0to30sec.
APNEA FLOW Breathing/ventilation has stopped. Patient must immediately be
ventilated manually!
Warning
(31)
>30 sec.
Check patient's spontaneous breathing
ability.
Check ventilator setting.
Insufficient fresh-gas supply. Check fresh-gas setting.
Tube kinked.
Leak in hose system.
Check hose system and tube.
Caution
(24)
0to30sec.
APNEA PRESSURE Breathing/ventilation has stopped. Patient must immediately be
ventilated manually!
Warning
(see
page 222)
(31)
≥0
or
>30 sec.
Insufficient fresh-gas supply. Check fresh-gas setting.
Leak or blockage in tube or hose system. Check hose system, tube, and microbial
filter.
Patient not connected. Connect patient correctly.
Caution/
Advisory
(see
page 213)
(11/9)
APNEA VENTILATION No spontaneous breathing efforts by the
patient during Pressure Support mode.
Check the patient's trigger capability.
Set an adequate trigger.
Caution
(13)
Advisory
(7)
BATTERY LOW The battery capacity (Advisory =
10 to 20 %; Caution = <10%) of the
uninterruptible power supply is almost
exhausted.
Connect to mains power.
Check patient's condition!
Prepare manual ventilation with
100% O
2.
Warning
(26)
BREATH. SYS. TEMP. HIGH Breathing system temperature is too high. Check breathing system and breathing
gas temperatures.
Call DrägerService.
Priority Message Cause Remedy