Troubleshooting
Instructions for Use Apollo SW 4.5n 271
Part Number: 9053586, 3rd edition
Advisory
(6)
FAN FAIL Fan for evacuating gases inside the
device is defective.
Anesthesia machine must be
switched off as quickly as possible!
Defective fans in combination with an
internal leakage may lead to elevated O
2
concentrations inside the anesthesia
machine.
Risk of fire!
Call DrägerService.
Caution
(14)
FG FLOW TOO HIGH Total fresh-gas flow is above 19 L/min. Reduce fresh-gas flow.
Check vaporizer setting.
Caution
(16)
Warning
(31)
FG LOW OR LEAK Fresh-gas setting too low. The priority of
the warning depends on the extend of the
fresh-gas shortage.
Increase fresh-gas flow.
Check anesthetic gas receiving system
AGS.
Leak. Repair leak.
Advisory
(1)
GAS SENSOR FAIL
(in Standby only)
Complete gas measurement system
failure.
Use external gas measuring system.
Call DrägerService.
Warning/
Advisory
(29/10)
GAS MIXER FAIL* Fresh-gas measurement is probably
inaccurate or failed.
Switch over to Aux CGO may have failed.
Check setting of vaporizer unit.
Use only oxygen as fresh gas and check
total flow meter.
Call DrägerService.
Warning/
Advisory
(30/10)
GAS + VENT FAIL.* Ventilator failed.
Fresh-gas measurement is probably
inaccurate or failed.
Switch over to Aux CGO may have failed.
Patient must immediately be
ventilated manually!
Use only oxygen as fresh gas and check
total flow meter.
Check setting of vaporizer unit.
Call DrägerService.
Warning
(27)
HIGH AIRWAY PRESSURE Upper alarm limit for the airway pressure
has been exceeded.
Ventilation hose kinked. Check hose system and tube.
Stenosis.
Ventilation settings are not correct. Correct the ventilation settings.
Caution/
Advisory
(14/10)
INCORRECT FG FLOW Set fresh-gas flow cannot be delivered. Reduce fresh-gas flow for each gas
below 12 L/min.
Check total flow meter.
Call DrägerService.
Caution
(11)
INSP CO
2 HIGH Soda lime in circle system exhausted. Increase fresh-gas flow.
Replace soda lime.
Leak or fault in breathing system. Replace breathing system.
High respiratory rates.
At high ventilation frequencies, the
measured value cannot keep up
completely with the gas concentration for
reasons due to the system.
Adjust alarm limits if necessary.
Dead space ventilation. Check ventilation settings.
Priority Message Cause Remedy

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals