Troubleshooting
272 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Advisory
(8)
INSP. FLOW SENSOR FAIL Inspiratory flow sensor is defective. Replace flow sensor (see page 231).
Advisory
(10)
Caution
(24)
Warning
(31)
INSP. HAL. HIGH
INSP. ISO. HIGH
INSP. ENF. HIGH
INSP. DES. HIGH
INSP. SEV. HIGH
Caution (24) =
insp. MAC value >3 MAC
for >180 seconds.
Warning (31) =
insp. MAC value >5 MAC
Warning (31) =
insp. MAC value >3 MAC and
exp. MAC value >2.5 MAC
for >30 seconds.
Inspiratory anesthetic gas concentration
exceeds 5 MAC.
Check vaporizer and fresh-gas settings.
Inspiratory anesthetic gas concentration
exceeds 3 MAC for more than
180 seconds.
Inspiratory anesthetic gas concentration
exceeds 3 MAC and the expiratory
concentration exceeds 2.5 MAC for more
than 30 seconds.
Advisory (10) =
insp. gas concentration > upper alarm
limit for 0to30seconds.
(preliminary information for
the user)
Caution (24) =
insp. gas concentration > upper alarm
limit for 31 to 180 seconds.
Warning (31) =
insp. gas concentration > upper alarm
limit for >180 seconds.
Inspiratory anesthetic gas concentration
exceeds the high alarm limit for at least
two breaths.
Check vaporizer and fresh-gas settings.
Priority Message Cause Remedy