System Setup
60 Instructions for Use Apollo SW 4.5n
Part Number: 9053586, 3rd edition
Figure 16. Location of AC Power LEDThe battery must be charged for 10 hours before
using the anesthesia machine for the first time:
1. Plug the mains power plug of the Apollo
anesthesia machine into the mains outlet.
The mains voltage must correspond to that specified
on the rating plate on the back of the machine.
The green LED labeled >
N AC Power< lights up
(1 in Figure 16).
2. Leave the Apollo connected to the mains for 10
hours. The anesthesia machine does not have to
be switched on.
WARNING!
Risk of device failure
If the batteries have not been sufficiently
charged and a power failure occurs, it may only
be possible to continue operation for a short
period of time.
Charge batteries for at least 10 hours before
first use or after storage.
CAUTION!
Risk of device failure
In the event of a power failure, any devices
connected to auxiliary power outlets will not be
powered by the UPS.
Pay special attention to all power indicators of
connected devices.
CAUTION!
Risk of electric shock and of device malfunction
There is a risk of injury to the user or damage to the
device if the device is connected to a power socket
with the wrong mains voltage or without a protective
conductor.
The power cable must only be connected to a
power socket with a protective conductor, see
“Specifications”.
WARNING!
Risk of battery failure
Allowing the battery to run low can damage it.
It must be charged at least every four weeks.