The terminal configurations have been set by the manufacturer depending on the specifications ordered. For individual
wiring of additional components, see “5.1.1. Pin Assignments” on page20. To open the terminal compartment, unscrew the
enclosure cover.
A Power supply (AC/DC)
B Connectors I/Os, relay outputs
C Sensor connection block
D MODBUS (RS-485, RS-232)
E LAN, RJ45, SIM Card
F Fuse
G Cable entries blocks
H LTE/4G antenna socket
I WiFi antenna socket
Figure 7: Overview terminal compartment
OTE: N The 4G/LTE dipole antenna is connected via the antenna connector on the enclosure bottom next to the
cable entry blocks.
MPORTANTI
Specification for cables for wiring:
Use copper cable only
Temperature: min. 158° F (70° C)
Cable diameter: 0.04 in. (1 mm
Electrical Connection
Page 19 March 2021 HYB-UM-03155-EN-03