7.12.2. Date Time
Settings for time and date and option for time synchronization via internet.
Time synchronization via NTP server can be activated to synchronize the system time of the measuring system or PC with
other data bases or networks to enable identical time stamps.
MPORTANTI
Save all measurement data (see chapter 7.6.1) and log files (see chapter 7.11.1) before changing system time or date. Otherwise,
data could be deleted or overwritten. Measurement data and log files will be deleted if the last change to a file is too long in the past.
The accepted time limit is 12 months for measurement data and 90 days for log files.
Date Time
Setting Description
Current time
System time > Display of current system time and date.
Date Time
Time manual Check radio button to set time and date manually.
After checking the button, the current time and date are displayed.
Click on the respective input field to change the time and date
Time > Set desired time
Date > Enter desired date
Click Apply to save entries.
Internet time Check radio button to activate time and date setting via NTP server (Network Time Protocol)
Server.
NTP uses UTC (Coordinated Universal Time) for synchronization. In Germany MEZ (= CET)
and MESZ (= CEST) is applied; CET = UTC + 1h respectively CEST = UTC + 2h
Programming
Page 95 March 2021 HYB-UM-03155-EN-03