7.9.4. Totalizer
Totalizer function for measurement and display of total and daily volume values (V).
Flow is measured in either direction (bi-directional):
Volume positive +V (in flow direction), Volume negative -V (against flow direction)
Set, define or change required unit for Volume under Settings Units.
Totalizer
Setting Description
Total volume
Total Volume Internal sum counter of total volume
> Display of volume value measured since first system start / first commissioning. This value
cannot be reset to zero.
Daily volume from YYYY/MM/DD
Daily volume Daily volume counter (V difference)
> Display of daily volume value
+/- sign are taken into account in summation of V
V (daily volume) = +V (volume positive) -V (volume negative)
Daily volume
absolute
> Display absolute volume per day (day total)
(- sign of negative volume is not taken into account!)
V (daily volume absolute) =
V (volume positive) + V (amount volume negative)
Daily volume
positive
Positive counter (+ V):
> Display of daily volume against flow direction
Daily volume
negative
Negative counter (-V):
> Display of daily volume in flow direction
Volume since last reset
Volume Volume counter (V difference)
> Display of volume value
+/- sign are taken into account in summation of V
V (daily volume) = +V (volume positive) -V (volume negative)
Programming