Water level
Setting Description
Switching (continued) There are two possible situations depending of the direction to which the water level changes.
Situation A
The water level is higher than value of Switching level + (0.5 × Hysteresis).
Water level of additional water level sensor will be used.
Not before the water level falls below the value Switching level – (0.5 × Hysteresis)
software is switching automatically, and water level of area velocity sensor will be used.
Situation B
The water level is lower than value Switching level – (0.5 × Hysteresis)).
Water level of area velocity sensor will be used.
Not before the water level rises above the value of Switching value + (0.5 × Hysteresis)
software is switching automatically, and water level of additional water level sensor will be used.
Click Apply to save settings.