Commissioning Proline t-mass I 300 HART
94 Endress+Hauser
Minimum frequency value
→ 94
Maximum frequency value
→ 95
Measuring value at minimum
frequency
→ 95
Measuring value at maximum
frequency
→ 95
Failure mode
→ 95
Failure frequency
→ 95
Invert output signal
→ 95
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Operating mode – Define the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
–
Terminal number – Shows the terminal numbers
used by the PFS output
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Signal mode – Select the signal mode for the
PFS output.
• Passive
• Active
*
• Passive NAMUR
–
Assign frequency output The Frequency option is
selected in the Operating
mode parameter (→ 92).
Select process variable for
frequency output.
• Off
• Temperature
• Mass flow
• Corrected volume
flow
• FAD volume flow
*
• Volume flow
• Energy flow
*
• Heat flow
*
• Density
• Flow velocity
• Pressure
• 2nd temperature
delta heat
*
• Electronic
temperature
–
Minimum frequency value The Frequency option is
selected in the Operating
mode parameter (→ 92)
and a process variable is
selected in the Assign
frequency output parameter
(→ 94).
Enter minimum frequency. 0.0 to 10 000.0 Hz –