PROline Prosonic Flow 93 FOUNDATION Fieldbus 3 Transducer Blocks (transmission blocks)
Endress+Hauser 111
System Value -
Sound Velocity
read only Displays the current sound velocity. The sound velocity is made
available to the downstream Analog Input function blocks as a
process variable.
Note!
The unit is selected and displayed in the “System Unit - Sound
Velocity” parameter.
System Unit -
Sound Velocity
AUTO - OOS For selecting the unit for the sound velocity.
Options:
m/s
ft/s
Factory setting:
m/s
System Value -
Flow Velocity
read only Displays the current flow velocity. The flow velocity is made
available to the downstream Analog Input function blocks as a
process variable.
Note!
The unit is selected and displayed in the “System Unit - Flow
Velocity” parameter.
System Unit -
Flow Velocity
AUTO - OOS For selecting the unit for the flow velocity.
Options:
m/s
ft/s
Factory setting:
m/s
System Value -
Signal Strength
read only Displays the current signal strength. The signal strength is made
available to the downstream Analog Input function blocks as a
process variable.
System Unit -
Temperature
AUTO - OOS For selecting the unit for the fluid temperature.
Options:
°C (Celsius)
K (Kelvin)
°F (Fahrenheit)
R (Rankine)
Factory setting:
Depends on country (°C or °F),
see factory setting Page 76 ff.
“TRANSDUCER_CH1” and “TRANSDUCER_CH2” (E+H parameters) /
Base index 1200 and 1300
Parameter Write access
with operating
mode
(MODE_BLK)
Description