5 Discrete Output function block PROline Prosonic Flow 93 FOUNDATION Fieldbus
190 Endress+Hauser
STRATEGY AUTO - MAN -
OOS - CAS -
RCAS
Parameter for grouping and thus faster evaluation of blocks.
Grouping is carried out by entering the same numerical value in
the STRATEGY parameter of each individual block.
Factory setting:
0
Note!
These data are neither checked nor processed by the Discrete
Output function block.
TAG_DESC AUTO - MAN -
OOS - CAS -
RCAS
Entry of a user-specific text of max. 32 characters for unique
identification and assignment of the block.
Factory setting:
"_ _ _ _ _ _ _ _" no text
UPDATE_EVT read only Displays whether static block data have been modified, including
date and time.
XD_STATE AUTO - MAN -
OOS - CAS -
RCAS
In this parameter, the fieldbus host system can store a reference
to a description of the possible states of the PV_D parameter.
Note!
This parameter is not evaluated by PROline Prosonic Flow 93
FOUNDATION Fieldbus.
Discrete Output function block / base index 780