J22 TDLAS Gas Analyzer 4900002277
SpectraSensors, Inc. 63
Signal mode Select the signal mode
for the S output.
Passive
•Active
Passive
Switch output
function
In the Operating
mode parameter
the Switch option is
selected.
Select function for
switch output.
•Off
•On
Diagnostic
behavior
•Limit
Diagnostic behavior
Assign diagnostic
behavior
•In the Operat-
ing mode
parameter, the
Switch option
is selected.
•In the Switch
output func-
tion parame-
ter, the
Diagnostic
behavior
option is
selected.
Select diagnostic
behavior for switch
output.
•Alarm
•Alarm or
warning
Warning
Alarm
Assign limit In the Operat-
ing mode
parameter, the
Switch option
is selected.
•In the Switch
output func-
tion parame-
ter, the Limit
option is
selected.
Select process variable
for limit function.
Concentration Off
Assign status The Switch
option is
selected in the
Operating
mode parame-
ter.
•The Status
option is
selected in the
Switch output
function
parameter.
Select the device status
for switch output.
—Off
Parameter Prerequisite Description User Entry Factory Setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors J22 TDLAS and is the answer not in the manual?

Endress+Hauser SpectraSensors J22 TDLAS Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors J22 TDLAS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals