4900002277 J22 TDLAS Gas Analyzer
64 SpectraSensors, Inc.
Switch-on value In the Operat-
ing mode
parameter, the
Switch option
is selected.
•In the Switch
output func-
tion parame-
ter, the Limit
option is
selected.
Enter measured value
for the switch-on
point.
Signed floating-
point number
0 ppmv
Switch-off value In the Operat-
ing mode
parameter, the
Switch option
is selected.
•In the Switch
output func-
tion parame-
ter, the Limit
option is
selected.
Enter measured value
for the switch-off
point.
Signed floating-
point number
0 ppmv
Switch-on delay The Switch
option is
selected in the
Operating
mode parame-
ter.
•The Limit
option is
selected in the
Switch output
function
parameter.
Define delay for the
switch-on of status
output.
0.0 to 100.0 s 0.0 s
Switch-off delay The Switch
option is
selected in the
Operating
mode parame-
ter.
•The Limit
option is
selected in the
Switch output
function
parameter.
Define delay for the
switch-off of status
output.
0.0 to 100.0 s 0.0 s
Failure mode Define output behavior
in alarm condition.
Actual status
•Open
•Closed
Open
Invert output
signal
Invert the output
signal.
•No
•Yes
No
Parameter Prerequisite Description User Entry Factory Setting