9  Measurement
9.2  Measurement settings FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
100
Measuring range
After an analog output has been activated in the program branch Output Options, the
measuring range of the source item has to be entered.
 
• Select sign if the sign of the measured values is to be considered for the output.
• Select absolut if the sign of the measured values is not to be considered for the out-
put.
• Press ENTER.
• Enter the lowest expected measured value. The unit of measurement of the source item 
will be displayed.
Zero-Scale Val. is the value assigned to the value Output MIN of the output range.
• Press ENTER.
• Enter the highest expected measured value. The unit of measurement of the source 
item will be displayed.
Full-Scale Val. is the value assigned to the value Output MAX of the output range.
• Press ENTER.
Output Options\...\Meas.Values
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
output: current output
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA