Chapter 5. Communication
122 AT600 User’s Manual
Command 198 (0xC6): Read Fluid Information
This command is to read fluid information.
Command 200 (0xC8): Write Pipe Size
This command is to write pipe size.
Table 111: Request Data Bytes for Read Fluid Information
Byte Format Description
None
Table 112: Response Data Bytes for Read Fluid Information
Byte Format Description
0 - 3 Unsigned-32 Fluid Type:
0: Other
1. Water
4 - 7 Float Fluid SOS
8 – 11 Float Fluid minimum SOS
12 - 15 Float Fluid Maximum SOS
16 - 19 Float Fluid Temperature
Table 113: Command-Specific Response Codes for Read Fluid Information
Code Class Description
0 Success No Command-Specific Errors
1-5 Undefined
6 Error Device-Specific Command Error
7-127 Undefined
Table 114: Request Data Bytes for Write Pipe Size
Byte Format Description
0 Unsigned-8 Pipe size unit
1 - 4 Float Pipe OD Value
5 - 8 Float Pipe ID Value
9 - 12 Float Pipe WT Value

Table of Contents