EasyManua.ls Logo

GE AT600 - Page 144

GE AT600
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5. Communication
132 AT600 User’s Manual
Command 210 (0xD2): Read Reynolds KFactor Table
This command is to read Reynolds KFactor table.
Table 141: Request Data Bytes for Read Reynolds KFactor Table
Byte Format Description
0 Unsigned-8 Reynolds KFactor Index ( 1 - 6 )
Table 142: Response Data Bytes for Read Reynolds KFactor Table
Byte Format Description
0 Unsigned-8 Reynolds KFactor Index ( 1 - 6 )
1 – 4 Float Reynolds Value
5 – 8 Float Reynolds KV Value;
Table 143: Command-Specific Response Codes for Read Reynolds KFactor Table
Code Class Description
0 Success No Command-Specific Errors
1 Undefined
2 Error Invalid Selection
3-4 Undefined
5 Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7-127 Undefined

Table of Contents