IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-38 Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-4 Receive Test (cont’d)
7) Check the test result message and proceed as follows:
If one or more channels were identified as bad, trackball to the Save Report button and press
Select, then contact a GE Medical Systems Service Representative.
If no channels were identified as bad, mark the channels suspected as being bad as follows:
Trackball to each of the channels in the Display area and press select.
Trackball to the Mark bad channel checkbox and press Select.
Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.
8.) Trackball to the Low Freq Test button and press Select. The receive test is performed in low
frequency mode, and the test results are displayed in the Report dialog box, as shown in Figure 7-
20 on page 7-36: