IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-39
7-3-8-4 Receive Test (cont’d)
The Report dialog box contains the following information:
• Test Result Message: Displays the test result.
• TR4 Modules: Displays a graphic representation of the front board (four groups of TR4 sockets).
Each group consists of four TR4 sockets, each socket containing four vectors (channels), as
follows:
- Each vector is represented by a button (highlighted in green) that displays the number of the
socket to which the TR4 is connected.
- Each button is highlighted yellow when selected.
- Each button is highlighted red if errors were detected for one or more channels.
• Checkboxes: Enable you to manipulate bad channels:
- Show Only Bad Channels: Select this checkbox to display only bad channels in the Display
area.
- Mark bad channel: Select this checkbox to manually mark the currently selected channel as
bad.
Figure 7-21 Report Dialog Box - Receive Test Low Frequency