7-40 Section 7-3 - Performing Front End (FE) Diagnostics
7-3-8-4 Receive Test (cont’d)
9.) Check the test result message and proceed as follows:
• If one or more channels were identified as bad, trackball to the Save Report button and press Select,
then contact a GE Medical Systems Service Representative.
• If no channels were identified as bad, mark the channels suspected as being bad as follows:
- Trackball to each of the channels in the Display area and press select.
- Trackball to the Mark bad channel checkbox and press Select.
- Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.
• If no errors were identified, proceed with the next test.
• If no errors were identified but the problem persists, contact your local GE Medical Systems field
engineer for assistance.
• If errors were identified but the results were marginal, perform calibration, as described in the
Calibration section, on page 6-14, and then repeat the test.
• If errors were identified and the results were not marginal, verify that all image port and PCI bridge
cables are connected and functioning properly, as described in Chapter 5 - Components and Function
(Theory). If the problem persists, contact your local GE Medical Systems field engineer for
assistance.
10.)Trackball to the Exit button and press Select. If required, repeat this test with a different pulser.