IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-49
7-3-8-8 RAM Test (cont’d)
Figure 7-26 F.B. Tab - FE Diagnostics Dialog Box - RAM Test (RFT Configuration)
3) Check the test results and proceed as follows:
If no errors were identified, proceed with the next test.
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page 6-21, and then repeat the test.
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in Chapter 5 - Components
and Function (Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
4) Reboot the system.