GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-49
7-3-8-8 RAM Test (cont’d)
Figure 7-26 F.B. Tab - FE Diagnostics Dialog Box - RAM Test (RFT Configuration)
3) Check the test results and proceed as follows:
• If no errors were identified, proceed with the next test.
• If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
• If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page 6-21, and then repeat the test.
• If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in Chapter 5 - Components
and Function (Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
4) Reboot the system.