GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-64 Section 7-3 - Performing Front End (FE) Diagnostics
7-3-9-6-2 AC Power Input Test (cont’d)
4) Trackball to the AC Power Input button and press Select. The AC Power Input dialog box is
displayed:
Figure 7-38 AC Power Input Dialog Box
5) Trackball to the Show button and press Select to display the voltage.
NOTE: For repeat readings, select Loop every and set the reading time cycle.