GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-85
8-5-12-5 BEP2 - ECG Module Removal Procedure (cont’d)
9) Disconnect the footswitch cable from the KB Controller card.
10.)Disconnect the COM2 cable from the rear of the ECG Module.
11) Carefully slide out the ECG Module from the front of the BEP.
Figure 8-72 ECG Module Removal- BEP2
Two screws